Home rTMS Challenges #1: Power Requirements

Why a treatment that lives in a hospital-grade cart isn't automatically ready to live on your kitchen counter, and how the engineering actually gets solved.

September 10, 2026 · 9 min read

If you've ever stood next to a clinical TMS machine, the first thing you notice isn't the coil or the chair. It's the cart. It's the size of a small refrigerator, it hums, and somewhere in the back there's a cable thick enough to belong to a washing machine. That's not an accident of industrial design. It's a direct consequence of physics, and it's the first real engineering wall that any company trying to bring repetitive transcranial magnetic stimulation (rTMS) into someone's home has to climb over.

This is the first post in a series on the practical challenges of home rTMS. We're starting with power, because it's foundational. Get it wrong, and nothing else about a home device (its size, its cost, its safety) works out the way you'd want.

Why TMS needs so much power in the first place

TMS works by passing a very brief, very large pulse of current through a coil held near the scalp. That current creates a rapidly changing magnetic field, and, through the same principle behind an electric transformer, that changing magnetic field induces its own electrical current inside the brain tissue underneath it, strong enough to depolarize neurons non-invasively, through the skull, without a single incision. That's the whole trick of TMS: magnetic fields pass through bone and skin essentially unimpeded, so you can induce current inside the head using a coil sitting outside it.

The catch is in the word "rapidly." The induced electric field in the tissue is proportional not to the current itself, but to how fast that current changes: its di/dt. To get a field strong enough to fire neurons, you need the current in the coil to swing from zero to several thousand amps and back again in well under a millisecond. A standard overview from a US National Institute of Mental Health webinar (presented by Dr. Zhi-De Deng) puts real numbers on this: commercial stimulators charge a capacitor to 2 kilo-volts then discharge it into the coil to produce peak currents in the coil of several thousand amps, with the whole pulse rising in about 100 microseconds and fields reaching up to 2 Tesla ([NIMH](https://www.nimh.nih.gov/)).

There is no wall outlet on Earth that can deliver several thousand amps on demand, which is exactly why nobody tries to.

The trick: don't draw the power, store it

This is the part that surprises people who aren't engineers: a TMS device doesn't pull its stimulation current from the wall in real time at all. It pulls a comparatively modest, steady trickle of power from the wall, banks it in a capacitor over the better part of a second, and then dumps the entire stored charge into the coil almost instantaneously. The wall outlet fills the bucket. The capacitor is the bucket. The coil pulse is the bucket tipped over in a fraction of a millisecond.

The physics has a clean shorthand for this. A capacitor's stored energy is E = ½CV² (half the capacitance times the voltage squared), and once that switch (historically a thyristor, increasingly an IGBT in newer designs) fires, essentially all of it leaves the capacitor and enters the coil within roughly 100 microseconds. Do the arithmetic on a few hundred joules in 100 microseconds and the peak power moving through that circuit for an instant is on the order of a few megawatts, comparable to the output of a small power plant, for a sliver of time shorter than a camera flash. The average power draw from your wall, over the whole second it took to recharge the capacitor for the next pulse, might be a few hundred watts, not far off from a hair dryer. The capacitor is what bridges that roughly ten-thousand-fold gap between what a wall outlet can sustain and what a coil needs in the instant it fires.

Recent circuit-design research shows this isn't just textbook theory. It's an active area of electrical engineering. A 2024 study in Sensors on TMS pulse-circuit design describes stimulators charging capacitor banks to voltages like 1,800V and 1,200V and discharging them through the coil in 30 to 100 microsecond pulses (Wang et al., Sensors 2024). A companion line of work in Frontiers in Neuroscience has focused specifically on redesigning that discharge waveform to cut the energy wasted as heat: one optimized design cut ohmic losses per pulse from 4.42 joules down to 2.58 joules, a 40% reduction, purely through smarter pulse shaping (Frontiers in Neuroscience, 2024). None of this is cosmetic. Every joule saved per pulse is a joule the power supply, the capacitor bank, and the cooling system don't have to be built to handle, which matters enormously once you're trying to fit a stimulator into something smaller than a hospital cart. This lineage of engineering traces back to Barker, Jalinous, and Freeston's original 1985 demonstration that a capacitor-discharge circuit could stimulate the human motor cortex non-invasively in the first place, the paper that started the field (The Lancet, 1985).

That capacitor-and-switch design is also why the mains-power question is really two separate questions that are easy to conflate: how much current the coil needs at the instant of the pulse (answer: always huge, capacitor or no capacitor), and how much continuous power the wall has to supply to keep recharging that capacitor at a clinically useful pace, say, 10 pulses a second for a full 3,000-pulse protocol. It's that second number that a home device's electrical design actually has to solve for, and it's where real product decisions, and real differences between devices, show up.

What's actually available in a US home

To know whether a device fits in a house, you have to know what a house actually offers. Standard US residential wiring runs at 120V, and the National Electrical Code caps most general-purpose circuits at either 15 or 20 amps. Do the multiplication and a 15A/120V circuit tops out at 1,800 watts; a 20A circuit at 2,400 watts. But NEC also has a "continuous load" rule: any circuit expected to run at or near full draw for three hours or more may only be loaded to 80% of its rating, which brings the practical ceiling down to about 1,440 watts on a 15A circuit and 1,920 watts on a 20A circuit (NEC continuous-load explainer). Beyond that, a typical house also has a handful of 240V, 30 to 50A circuits, but those are hard-wired to specific large appliances (the dryer, the range, an EV charger) with their own dedicated outlet shapes, not something you can casually repurpose for a medical device in a bedroom.

So a device meant to plug into an unmodified American home outlet, with no electrician, no new circuit, and no adapter, realistically has a budget of somewhere under about 1,800 watts of continuous draw, and ideally comfortably under that, since a device sharing a circuit with lamps, a phone charger, and whatever else happens to be plugged into that room shouldn't be the thing that trips the breaker.

How that budget plays out across real devices

This is where "portable" and "home-compatible" turn out to be two different engineering problems that get talked about as if they're the same one. A device can be small and wheel-able and still be built around clinic-grade electronics that assume a very different kind of outlet than the one in your living room.

DeviceManufacturerPublished power specWhat it means for home use
MagPro X100MagVenture230V mains, 2,700VA max; can run on 100/120V "through transformer" (spec sheet)2,700VA at 120V works out to about 22.5A, beyond even a 20A US circuit. Designed around 230V, the standard in most of the world outside North America; needs a step-up transformer or dedicated setup to run in a typical US room at all. Cart-based, about 84 lbs.
MagVenture GoMagVentureNo 510(k) has been filed under the "Go" name. Built on the MagPro R20 platform (product announcement); R20's own manufacturer catalog lists 230V mains, 800VA max, "operation from 100V or 120V through transformer" (R20 spec sheet)800VA at 120V works out to only about 6.7A, comfortably inside even a 15A circuit once stepped up, versus the X100's 22.5A. The catch is the transformer: this is a 230V-native device that needs an added step-up transformer to run on US mains at all.
BrainsWay Deep TMSBrainsWay100 to 240VAC; fuse rated 100-120V/20A, 200-240V/16A (professional FAQ)The 20A fuse rating at 120V is the tell. Built to expect a dedicated 20A circuit, standard in a clinic but not guaranteed in a bedroom.
Magstim HorizonMagstim3kVA; standard is 230V, with 115V available "by contacting Customer Service" (2025 catalog)Built around 230V as the default. A US-voltage version exists but is treated as a special case, not the standard configuration.
NeuroStarNeuroneticsThird-party figure: MagVenture's own FDA filing lists NeuroStar's power source as 110Vac, 20A in its predicate comparison table (K251119), reported by a competitor rather than stated by Neuronetics itselfIf that figure holds, NeuroStar needs the same dedicated 20A circuit as BrainsWay and the MagPro family. Large console/cart clinic system, not marketed as a home device.
Blossom TMSSEBERS MedicalStandard 120V outlet, "plug-and-play," with no electrical modifications in a standard outpatient facility (company blog, September 2024)A clinic system, not a home device: SEBERS Medical says it runs from a standard 120V outlet in a standard outpatient facility with no electrical modifications, so no dedicated circuit and no transformer. The coil is oil-cooled rather than air-cooled, which removes the airflow path and much of the noise, a detail that matters more in a bedroom than in a clinic.
Ampa OneAmpa120V, 15A dedicated circuit (company product page)Built around the actual constraint: a single standard 120V/15A US household outlet, on its own dedicated circuit. Described as suitable for "residential settings...connected to the public low-voltage power supply network."
SPRY TMS Therapy SystemSoterix MedicalUniversal input, 100-240V, 50/60Hz (SPRY TMS); FDA-cleared Feb. 20, 2026 (510(k) K251653). Soterix Medical confirmed to us directly (September 2026) that the system draws 15AUniversal voltage with no step-up transformer needed, and a draw low enough for a standard 15A US household circuit, the same target Ampa One is publicly documented at.

A pattern falls out of that table pretty clearly. Most of the established clinical TMS fleet, MagVenture's cart systems (including the R20 platform underneath Go), BrainsWay, Magstim, and per the flagged third-party figure above, NeuroStar too, was engineered around 230V mains, the global default everywhere outside North America and the assumption baked into products designed by European manufacturers. A step-up transformer can bridge that gap, and R20's own low 800VA draw makes the math on that bridge much friendlier than it is for the power-hungry X100. Blossom TMS complicates the 230V story in a useful way: it's a clinic-cleared system (K220625) that SEBERS Medical says runs from standard 120V outlets in a standard outpatient facility with no electrical modifications at all, so the transformer question never comes up, though it is a clinic device and not positioned for the home. The devices explicitly aimed at the home, right now that's Ampa One and possibly, SPRY, solve this by designing to the constraint from day one rather than adapting a clinic product after the fact. Ampa One publishes the clearest documented number: a standard 120V/15A circuit, the kind already in nearly every American bedroom, with no electrician visit required. SPRY lands in the same place: a universal 100-240V input per its own product page, and a 15A draw that Soterix Medical confirmed to us directly, though we'd still like to see that second figure show up in published documentation. That's a genuinely different design target from a 230V clinic machine, not just a smaller box around the same electronics.

The deeper engineering tension: power vs. portability

It's worth being honest that this isn't a solved problem so much as an active trade-off, and the research literature is pretty candid about that trade-off. A 2024 Sensors paper on miniaturized rTMS coils built a genuinely portable, 12.6-gram coil design, but to get there, the authors had to drop their capacitor voltage and peak current to 350V and 3.1 kA, "substantially lower than commercial systems using 1,200-3,000V," specifically because they were constrained by the voltage ratings of affordable, small-form-factor power switches (Sensors, 2024). Energy per pulse in that design worked out to about 23 joules, versus the "several hundred joules" that Ingenia's overview cites for standard clinical stimulators. The authors also flagged heat as a real limiting factor: even at that reduced power, sustained high-frequency pulsing raised coil temperature and would need further thermal-management work to run for long clinical protocols.

That's the honest state of the field: shrinking the power supply and the wall-power footprint doesn't necessarily come for free. Every design has to answer, with real engineering rather than marketing language, how it preserves adequate stimulation depth and intensity while giving up the electrical overhead of a clinic-grade power supply. That's precisely the kind of problem a home rTMS device has to solve rigorously rather than gloss over.

The takeaway

The reason your clinic's TMS machine looks like hospital equipment isn't that anyone over-engineered it. It's that a coil pulse genuinely needs several thousand amps for a fraction of a millisecond, and the capacitor-discharge circuit that makes that possible (a design lineage running from Barker's original 1985 demonstration through today's optimized pulse-shaping research) has historically been built around the assumption of clinic-grade power: 230V mains, or a dedicated 20A circuit at minimum. A standard American bedroom outlet offers neither of those by default. It offers 120V and, realistically, well under 1,800 continuous watts.

Getting rTMS safely and reliably into a home isn't a matter of making the cart smaller. It's a matter of adapting the power electronics, the capacitor bank, the charging circuit, the coil itself, to the roughly 1,800-watt, 120V reality of an ordinary household outlet.

That's challenge #1. Next month: challenge #2.

Sources

  1. National Institute of Mental Health, "Zhi-De Deng: Transcranial Magnetic Stimulation: Physics, Devices, and Modeling", April 2020
  2. Barker, A.T., Jalinous, R., Freeston, I.L., "Non-invasive magnetic stimulation of human motor cortex," The Lancet, 1985. The original demonstration of capacitor-discharge TMS.
  3. Wang et al., "An Efficient Pulse Circuit Design for Magnetic Stimulation with Diversified Waveforms and Adjustable Parameters," Sensors, 2024.
  4. "A novel pulse-current waveform circuit for low-energy consumption and low-noise transcranial magnetic stimulation," Frontiers in Neuroscience, 2024.
  5. "Design and Validation of Miniaturized Repetitive Transcranial Magnetic Stimulation (rTMS) Head Coils," Sensors, 2024.
  6. MagVenture, MagPro X100 technical specifications (230V, 2,700VA); MagPro R20 technical specifications (230V, 800VA, the platform MagVenture Go is built on); MagVenture Go announcement; FDA 510(k) summaries K251119 (MagPro mobile console, MDD/anxiety/OCD, lists "Power Source: 110Vac, 20A"), K170114 and K172667 (MagPro R20's own FDA filings, which don't include a power table).
  7. BrainsWay, "Does Deep TMS require any special equipment or facilities?"
  8. Magstim, 2025 TMS Catalog.
  9. Ampa, product specifications.
  10. Soterix Medical, SPRY TMS product specifications (100-240V input); SPRY FDA clearance announcement; FDA, 510(k) K251653 record; 15A draw per direct communication with Soterix Medical, September 2026 (not yet in a published spec sheet).
  11. Builder's Calculator, NEC continuous-load rules for 15A/20A residential circuits.
  12. Neuronetics has not published a power spec for NeuroStar in any filing we could find: K083538 (2008 original summary), K230029 and K231926 (later filings), and BrainsWay's original submission using NeuroStar as predicate, K122288, also omit it. The 110Vac/20A figure in the table comes from a different company's filing, MagVenture's K251119, which lists it for NeuroStar as the predicate device.
  13. Ingenia (Royal Academy of Engineering), "Transcranial magnetic stimulation". Overview of TMS capacitor-discharge physics, voltage/current/energy figures.
  14. SEBERS Medical, "The Quiet Power of Innovation: 3 Key Features of the Blossom TMS Device," September 9, 2024, and the company's accompanying Blossom TMS key-features graphic. Blossom TMS (K220625) is described as running from standard 120-volt outlets with plug-and-play installation; the graphic depicts a NEMA 5-15 plug seated in a standard US outlet, and the device uses an oil-cooled coil. SEBERS Medical describes the device as running from a standard 120V outlet in a standard outpatient facility without electrical modifications.