For clinicians

The clinical detail behind home TMS.

Seizure risk, screening, dose setting and supervision: the parts of this program that belong in front of a prescriber. This page is written for clinicians and researchers, and the patient-facing site stays deliberately quiet about it.

~62%of rTMS-related seizures occur in the very first session
RMTevery dose is set against the patient's own motor threshold
1in-home clinician visit before any unsupervised session

Safety profile

Seizure risk

Seizure is the most consequential adverse event in rTMS, and it is concentrated at the start of a course. Hover does not remove that risk. It puts a clinician in the room at the moment the risk is highest, then removes the degrees of freedom that would let it drift upward at home.

62%

About 62% of rTMS-related seizures occur during a patient's very first session. In the Hover protocol it is performed under direct clinical supervision, with the dose measured against the patient's own motor threshold rather than a fixed protocol.

The first session carries the risk

A Hover clinician performs the initial session in the patient's home, in person. Resting motor threshold is measured, the dose is set against it, and the coil is locked into the headgear at the calibrated position. No unsupervised session happens before that visit.

Dose is anchored to the patient

Intensity is expressed as a percentage of the patient's own motor threshold rather than a population default, and it is held in firmware. Neither the patient nor the app can raise it between sessions.

Nothing escalates at home

Session length, pulse count and waveform are fixed at the in-home visit. If placement, wear time or adherence deviate from the prescribed protocol, the next session is not authorized until the prescriber reviews it.

Sleep deprivation, alcohol use or withdrawal, untreated intracranial pathology and changes to a patient's medication list all shift seizure threshold and are reviewed before a course starts and again at the in-home visit.

Screening

Who should not be treated at home.

These are checked before a course starts. A clinician takes a full history and medication review at the in-home visit, before any session is authorized.

A seizure disorder or epilepsy

A history of seizures, a diagnosed seizure disorder or epilepsy excludes a patient from our protocol, regardless of how well controlled it is.

Metal implants in or around the head

Aneurysm clips, cochlear implants, deep brain or spinal stimulators, shunts and other ferromagnetic hardware near the coil path are exclusions.

Medications that lower seizure threshold

Tramadol, theophylline, tricyclics, antipsychotics, stimulants and some other agents are reviewed against the prescribing literature before a course is authorized.

Anything the in-home visit surfaces

Recent head trauma, unexplained neurological findings, pregnancy, or a medication change since screening are all reasons to pause rather than proceed.

Protocol

What the device enforces.

The home protocol assumes the patient is the least reliable component in the loop and designs around that: a clinician sets the course, the hardware holds it, and the prescriber sees every session.

  • Supervised first session

    A clinician is physically present for session one, measures motor threshold and sets every parameter before the headset is handed over.

  • Parameters fixed in firmware

    Intensity, frequency, waveform and pulse count are prescriber-set and locked. There is nothing at home to adjust, and no way to extend or repeat a session.

  • Placement verified every session

    Phone-camera photogrammetry compares the headset against the position captured at the in-home visit. On a mismatch, no session code is released.

  • Single-use session authorization

    Each session needs a fresh code, and the hardware freezes once the prescribed dose is complete — so a course cannot quietly become more than it was prescribed.

  • Data returned to the prescriber

    Placement, duration and adherence data return to the clinician who set the parameters, and deviations are flagged before the next session.

Hover is an investigational device still in development. The safeguards described here are our intended protocol; they have not yet been reviewed by the FDA for unsupervised home use. Nothing on this page is medical advice or a substitute for clinical judgement in an individual case.