Home rTMS Challenges
Home rTMS Challenges #3: Accuracy
How precisely does a TMS coil really need to be placed, how precise are today's clinical tools in practice, and what does that mean for treatment at home?
October 6, 2026 · 9 min read
In the first two posts in this series, we looked at power and portability and found that both are more manageable than they first appear. This week we're tackling the question we hear most often about home rTMS: how do you make sure the coil is in the right place without a technician and a neuronavigation system in the room?
It's a fair question. TMS works by stimulating a specific patch of cortex, and if the coil is in the wrong spot, the wrong patch gets stimulated. But "the right place" turns out to be less of a pinpoint than people tend to imagine. The field has spent two decades debating where exactly the target is, the tools that measure coil position all carry their own error, and when researchers have compared simple scalp-based targeting with full MRI-guided neuronavigation head-to-head, the results have been closer than you might expect.
That doesn't mean accuracy is unimportant. It means the real requirements are worth understanding before assuming home rTMS can't meet them.
Where is "the target," anyway?
For depression, the standard rTMS target is the left dorsolateral prefrontal cortex (DLPFC). But the DLPFC is a large region, and over the years clinicians and researchers have proposed many different spots within it.
The oldest is the "5 cm rule": find the spot on the scalp that makes the right thumb twitch (the motor "hot spot"), then move the coil 5 cm forward. It's simple, needs no imaging, and was the method used in the pivotal trials that established rTMS for depression. Later work showed that it often lands too far back, so variants such as a 5.5 cm or 6 cm rule were adopted.
Other targets are defined by brain anatomy, scalp landmarks, or brain networks. In an influential 2012 paper, Fox and colleagues collected the coordinates of the commonly used DLPFC targets in standard MNI brain space and showed that targets with better clinical results tended to be more strongly anti-correlated with a deep region called the subgenual cingulate (Fox et al., Biological Psychiatry, 2012). From that, they proposed an "optimized" target, often called the Fox target. Here are several of the targets they analyzed, along with the straight-line distance from each one to the Fox target, which we calculated from the published coordinates:
| Target | MNI coordinates (x, y, z) | Distance to Fox target |
|---|---|---|
| 5 cm rule | (−41, 16, 54) | ≈ 40 mm |
| EEG F3 (the Beam-F3 target) | (−37, 26, 49) | ≈ 29 mm |
| Center of BA9 | (−36, 39, 43) | ≈ 18 mm |
| Fitzgerald target | (−46, 45, 38) | ≈ 15 mm |
| Center of BA46 | (−44, 40, 29) | ≈ 8 mm |
| Fox efficacy-based target | (−38, 44, 26) | 0 mm |
Coordinates from Fox et al. (2012). Distances are Euclidean distances in MNI space, computed by us.
That spread is striking. The 5 cm rule target and the Fox target are about 4 centimeters apart. Even the 5 cm rule and the F3 site, the two most common scalp-based targets in clinical practice, sit about a centimeter apart.
And yet rTMS works across this whole range. The 5 cm rule, the target furthest from Fox's optimum, is the method behind the trials that brought rTMS into routine clinical use. Fox's own analysis suggests that the more anterior targets probably do better on average, and that matters. But it's a gradual trend across centimeters, not a sharp cliff where a few millimeters separate a treatment that works from one that doesn't.
The physics points the same way. A 2023 analysis of coil misplacement found that shifting a figure-8 coil by about 10 to 12 mm along the handle direction lowered the electric field at the target by only about 5%, and side-to-side shifts of 10 to 15 mm lowered it by about 10% (Koehler, Kammer & Goetz, 2023). A TMS coil doesn't stimulate a pinpoint. It stimulates a region several centimeters across, with the strongest field in the middle.
The takeaway isn't that position doesn't matter. It's that the evidence supports getting into the right region and staying there consistently, rather than chasing sub-millimeter precision.
Neuronavigation: precise, but not error-free
Neuronavigation is the gold standard for placing a coil. An optical camera tracks reflective markers on the patient's head and on the coil, the patient's head is matched (or "registered") to an MRI or a template brain, and the operator sees on screen in real time where the coil is relative to the target.
It's a powerful tool, but it has its own error budget. A careful 2022 study in the Journal of Neural Engineering measured each step of the navigation chain and found total registration accuracies of about 2.2 to 3.6 mm and 1°. The biggest single source of error was the head-to-MRI registration step, followed by movement of the tracker attached to the patient's head. Once the step of estimating where the electric field actually peaks in the brain was included, the peak location could be off by 1.5 to 5.0 mm (Nieminen et al., Journal of Neural Engineering, 2022).
Those are excellent numbers, measured under careful laboratory conditions. But they make an important point: even the best clinical tool puts the coil within a few millimeters of the target, not exactly on it. Day-to-day differences in how a technician registers the head, how the head tracker sits, and how the patient moves add further variation.
And neuronavigation tells you where the coil is relative to whatever target you chose. If the target itself is uncertain (and as the table above shows, the field hasn't settled on one), millimeter-level navigation to a centimeter-level debate about where to aim is precision layered on top of uncertainty.
Holding the position: robots, arms, and vacuum pillows
Getting the coil into position is only half the job. It also has to stay there for the whole session, which for standard protocols can mean 20 to 40 minutes.
Clinics and research labs use several strategies:
- Head fixation. Treatment chairs with head and neck rests, chin rests, and in some research settings vacuum pillows (bead-filled cushions that mold around the head and then stiffen when the air is pumped out) all try to keep the head still relative to a coil mounted on a fixed arm.
- Continuous neuronavigation. The operator watches the navigation display throughout the session and corrects the coil whenever it drifts.
- Robots. Robotic coil holders combine optical tracking with a motorized arm that follows the patient's head. The Axilum Robotics TMS-Cobot, for example, received FDA clearance in 2019 and uses real-time optical tracking to adjust coil position and orientation automatically to compensate for head movement (Medical Design & Outsourcing). A preprint evaluating a TMS robot reported median errors of about 1.3 mm and 3.5° over a session, with the errors coming mainly from small shifts in the optical trackers themselves, not from the robot (bioRxiv preprint, 2019).
All of these approaches share a basic design assumption: the coil is attached to the room, and the head is free to move. So the head has to be immobilized, or the coil has to chase it.
There's another option: attach the coil to the head. Then the head can move freely, because the coil moves with it. That's the key design insight behind Hover's fitted headset, and we'll come back to it below.
The high-precision end: SAINT
At the most precise end of the spectrum is Stanford Neuromodulation Therapy (SNT, originally called SAINT). It pairs an accelerated schedule (ten iTBS sessions a day for five days, 90,000 pulses in total) with an individualized target found in each patient's resting-state functional MRI, using the same subgenual cingulate anti-correlation idea Fox described, and delivered with neuronavigation. In its randomized, double-blind trial, 79% of patients in the active group reached remission, compared with 13% on sham (Cole et al., American Journal of Psychiatry, 2022).
Those are remarkable results, and they're often taken as proof that precise targeting is what drives efficacy. But SNT changes several things at once: the target, the dose (far more pulses), the schedule (compressed into five days), and the waveform (iTBS). The trial wasn't designed to isolate how much each ingredient contributes. Other trials have tried to do exactly that.
What happens when trials compare targeting methods directly?
This is where the evidence gets interesting. Several randomized trials have compared simpler targeting methods with more sophisticated ones, and most have found smaller differences than you might expect:
- Neuronavigation vs. the 5 cm rule (2009). In a randomized trial of 51 patients with treatment-resistant depression, those treated at an MRI-neuronavigated target improved more at four weeks than those treated with the 5 cm rule (Fitzgerald et al., Neuropsychopharmacology, 2009). This is the trial most often cited in favor of neuronavigation.
- Neuronavigation vs. the 5 cm rule (2025). A double-blind, multicenter randomized trial of 105 patients found no difference: 31.8% responded with MRI-based neuronavigation versus 35.6% with the standard 5 cm method (p = 0.58). The authors concluded that the study "failed to reproduce previous findings supporting the use of neuro-navigation system to optimize rTMS efficacy," while noting limits such as a short 10-session course (Millet et al., PLOS ONE, 2025).
- Beam-F3 vs. 5.5 cm. A randomized trial comparing two scalp-based methods, Beam-F3 and a 5.5 cm rule, found similar effectiveness between them (Brain Stimulation, 2023).
- Connectivity-guided targeting vs. F3. A five-center, double-blind trial of 255 patients compared individualized, fMRI connectivity-guided iTBS against standard rTMS at the F3 site (both delivered with MRI neuronavigation). Both groups improved by about 8 points on the Hamilton depression scale at 8 weeks, with no significant difference between them (Morriss et al., Nature Medicine, 2024).
Put together, the picture is nuanced. Precise targeting may help, especially for the most individualized protocols. But across several well-run trials, simpler and more reproducible methods have performed in the same range as more sophisticated ones.
Getting accurate without neuronavigation
That's why a lot of recent work has gone into scalp-based methods that get close to MRI-defined targets without needing an MRI or a navigation system:
- Beam-F3. Introduced in 2009, Beam-F3 uses three simple head measurements (nasion to inion, tragus to tragus, and head circumference) to locate the F3 position of the EEG 10-20 system, a common DLPFC target, with a tape measure (Beam et al., Brain Stimulation, 2009). It has become one of the most widely used targeting methods in clinical practice.
- MNI2CPC. This probabilistic method maps a target defined in standard MNI brain coordinates to a location on the scalp expressed in cranial proportional coordinates (CPC), a head-size-independent scalp coordinate system. It lets a clinician aim for any published brain coordinate, not just F3, using scalp measurements (Brain Stimulation, 2024).
- MRI-free head modeling (Valter et al.). This approach starts from the same three Beam-F3-style measurements, but uses them to scale a standard MNI head model into an individualized head model for each patient (Valter et al., Transcranial Magnetic Stimulation, 2025; CCNY Neural Engineering Group summary). Disclosure: the paper's first author is Hover's founder.
A purely measurement-based method landing within about 5 mm of an MRI-defined target is in the same range as the few-millimeter error of a full neuronavigation chain, and both are far smaller than the centimeters separating the targets the field still debates.
Accuracy at home: set once, reproduce every day
So what does a home system actually need to get right? Two things, and they're very different:
- Finding the target, once. This is a clinical decision. It can be made in the clinic, by a clinician, using whatever method the protocol calls for: Beam-F3, the 5 cm rule, or full neuronavigation.
- Reproducing that position, every session. This is a consistency problem. It's where home treatment has to perform, and it's where the clinic's tools (arms, chairs, vacuum pillows, robots, and navigation cameras) are built around the assumption that the coil is attached to the room.
Hover is designed around separating those two jobs:
- A fixed headgear. The clinician fits a headset to the patient and locks the coil at the chosen target. Because the coil is attached to the headset and not to a stand, the patient's head can move naturally during treatment, and the coil moves with it. No head fixation or robot is needed to keep it in place.
- A smartphone app with computer vision. At home, before each session, the patient uses an app on their phone. The app uses the camera and computer vision to compare how the headset sits on the head today with how it sat during the clinician-supervised fitting.
- Augmented-reality guidance. If the headset isn't where it should be, the app shows the patient on screen how to adjust it, with AR overlays, until the position matches the clinic reference. Treatment proceeds only once it does.
Instead of re-deriving the target every day with neuronavigation, the system reproduces a position the clinician already verified. The problem changes from "where is the DLPFC?" to "is the headset where the clinician put it?", and a phone camera is well suited to answering that second question.
The takeaway
Accuracy matters in TMS, but the requirement is often misunderstood. The commonly used DLPFC targets are spread across several centimeters. Even neuronavigation carries a few millimeters of error. And several randomized trials comparing simple scalp-based targeting with MRI-guided neuronavigation have found comparable outcomes. What the evidence supports isn't sub-millimeter precision. It's reliably landing in the right region and staying consistent from one session to the next.
That's a standard a home system can meet: find the target once in the clinic, fix the coil to the head rather than the room, and use the camera in the patient's pocket to confirm the setup matches before every session.
That's challenge #3. Next week: #4.
Sources
- Fox, M.D., Buckner, R.L., White, M.P., Greicius, M.D., Pascual-Leone, A., "Efficacy of Transcranial Magnetic Stimulation Targets for Depression Is Related to Intrinsic Functional Connectivity with the Subgenual Cingulate," Biological Psychiatry, 2012.
- Koehler, M., Kammer, T., Goetz, S.M., "Quantitative Impact of Coil Misalignment and Misplacement in Transcranial Magnetic Stimulation," bioRxiv, 2023.
- Nieminen, A.E., et al., "Accuracy and precision of navigated transcranial magnetic stimulation," Journal of Neural Engineering, 2022.
- Medical Design & Outsourcing, "FDA clears Axilum Robotics' system for TMS positioning," 2019.
- Robotic TMS coil-positioning evaluation, bioRxiv preprint 10.1101/641456, 2019.
- Cole, E.J., et al., "Stanford Neuromodulation Therapy (SNT): A Double-Blind Randomized Controlled Trial," American Journal of Psychiatry, 2022.
- Fitzgerald, P.B., et al., "A Randomized Trial of rTMS Targeted with MRI Based Neuro-Navigation in Treatment-Resistant Depression," Neuropsychopharmacology, 2009.
- Millet, B., et al., "Repetitive Transcranial Magnetic Stimulation targeted with MRI based neuro-navigation in major depressive episode: a double-blind, multicenter randomized controlled trial," PLOS ONE, 2025.
- A randomized trial comparing beam F3 and 5.5 cm targeting in rTMS treatment of depression demonstrates similar effectiveness," Brain Stimulation, 2023.
- Morriss, R., et al., "Connectivity-guided intermittent theta burst versus repetitive transcranial magnetic stimulation for treatment-resistant depression: a randomized controlled trial," Nature Medicine, 2024.
- Beam, W., Borckardt, J.J., Reeves, S.T., George, M.S., "An efficient and accurate new method for locating the F3 position for prefrontal TMS applications," Brain Stimulation, 2009.
- MNI2CPC: A probabilistic cortex-to-scalp mapping for non-invasive brain stimulation targeting," Brain Stimulation, 2024.
- Valter, Y., Truong, D.Q., Kawasumi, Y., Bikson, M., Datta, A., "TMS targeting using MRI-free head modeling," Transcranial Magnetic Stimulation, 2025; free calculator at tmstargets.com.